Computed tomography has greatly improved over the last decade, especially through x-ray dose exposure reduction while maintaining image quality. Herein, a new concept is proposed to improve the contrast-to-noise ratio (CNR) by including the time-of-flight (TOF) information of individual photons to obtain further insight on the photon's trajectory and to reject scatter contribution. The proof of the concept relies on both simulation and experimental measurements in a cone-beam computed tomography arrangement. Results show a statistical difference between the TOF of scattered and primary photons exploitable in TOF computed tomography. For a large volume of the size of a human abdomen, a scatter reduction from 296% to 4% is achieved in our simulation setup with perfect timing measurements which yields a 110% better CNR, or a dose reduction by a factor of four. Cup artifacts are also reduced from 24.7% to 0.8%, and attenuation inaccuracies are improved from -26.3% to -0.8%. With 100 ps and 10 ps FWHM timing jitters, respectively 75% and 95% of the scatter contribution can be removed with marginal gains below 10 ps. Experimental measurements confirm the feasibility of measuring statistical differences between the TOF of scattered and primary photons.

Rossignol, J., Martinez Turtos, R., Gundacker, S., Gaudreault, D., Auffray, E., Lecoq, P., et al. (2020). Time-of-flight computed tomography - proof of principle. PHYSICS IN MEDICINE AND BIOLOGY, 65(8) [10.1088/1361-6560/ab78bf].

Time-of-flight computed tomography - proof of principle

Martinez Turtos R.;Gundacker S.;
2020

Abstract

Computed tomography has greatly improved over the last decade, especially through x-ray dose exposure reduction while maintaining image quality. Herein, a new concept is proposed to improve the contrast-to-noise ratio (CNR) by including the time-of-flight (TOF) information of individual photons to obtain further insight on the photon's trajectory and to reject scatter contribution. The proof of the concept relies on both simulation and experimental measurements in a cone-beam computed tomography arrangement. Results show a statistical difference between the TOF of scattered and primary photons exploitable in TOF computed tomography. For a large volume of the size of a human abdomen, a scatter reduction from 296% to 4% is achieved in our simulation setup with perfect timing measurements which yields a 110% better CNR, or a dose reduction by a factor of four. Cup artifacts are also reduced from 24.7% to 0.8%, and attenuation inaccuracies are improved from -26.3% to -0.8%. With 100 ps and 10 ps FWHM timing jitters, respectively 75% and 95% of the scatter contribution can be removed with marginal gains below 10 ps. Experimental measurements confirm the feasibility of measuring statistical differences between the TOF of scattered and primary photons.
Articolo in rivista - Articolo scientifico
Algorithms; Artifacts; Cone-Beam Computed Tomography; Humans; Image Processing, Computer-Assisted; Phantoms, Imaging; Photons; Scattering, Radiation; Time Factors
English
2020
65
8
085013
none
Rossignol, J., Martinez Turtos, R., Gundacker, S., Gaudreault, D., Auffray, E., Lecoq, P., et al. (2020). Time-of-flight computed tomography - proof of principle. PHYSICS IN MEDICINE AND BIOLOGY, 65(8) [10.1088/1361-6560/ab78bf].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/527756
Citazioni
  • Scopus 23
  • ???jsp.display-item.citation.isi??? 18
Social impact