Silicon carbide detectors represent an alternative to diamond detectors for fast neutron detection in harsh environments, especially fusion plasmas. Previous studies on thin prototypes (either 10 μm or 100 μm thick) suggested that thicker active volumes might be better suited for spectroscopy measurements, due to the higher chance of retaining the neutron interaction products inside the active volume. Therefore, in this work two 250 μm SiC prototypes are tested with alpha particles following the same process conducted in the past for thinner prototypes. A stable detection is demonstrated, along an energy resolution that, if projected to DT neutrons, could become the lowest achieved so far with a SiC detector (1.3%). Some difficulties in reaching a full depletion are highlighted, as long as perspectives of a partial polarization operation of the detectors.

Kushoro, M., Rebai, M., La Via, F., Meli, A., Meda, L., Parisi, M., et al. (2023). Performance of a thick 250 μm silicon carbide detector: stability and energy resolution. JOURNAL OF INSTRUMENTATION, 18(3) [10.1088/1748-0221/18/03/C03007].

Performance of a thick 250 μm silicon carbide detector: stability and energy resolution

Kushoro, MH
;
Rebai, M;Perelli Cippo, E;Putignano, O;Tardocchi, M
2023

Abstract

Silicon carbide detectors represent an alternative to diamond detectors for fast neutron detection in harsh environments, especially fusion plasmas. Previous studies on thin prototypes (either 10 μm or 100 μm thick) suggested that thicker active volumes might be better suited for spectroscopy measurements, due to the higher chance of retaining the neutron interaction products inside the active volume. Therefore, in this work two 250 μm SiC prototypes are tested with alpha particles following the same process conducted in the past for thinner prototypes. A stable detection is demonstrated, along an energy resolution that, if projected to DT neutrons, could become the lowest achieved so far with a SiC detector (1.3%). Some difficulties in reaching a full depletion are highlighted, as long as perspectives of a partial polarization operation of the detectors.
Articolo in rivista - Articolo scientifico
Neutron detectors (cold, thermal, fast neutrons); Nuclear instruments and methods for hot plasma diagnostics; Solid state detectors;
English
2023
18
3
C03007
open
Kushoro, M., Rebai, M., La Via, F., Meli, A., Meda, L., Parisi, M., et al. (2023). Performance of a thick 250 μm silicon carbide detector: stability and energy resolution. JOURNAL OF INSTRUMENTATION, 18(3) [10.1088/1748-0221/18/03/C03007].
File in questo prodotto:
File Dimensione Formato  
Kushoro-2023-JINST-VoR.pdf

accesso aperto

Descrizione: Article
Tipologia di allegato: Publisher’s Version (Version of Record, VoR)
Licenza: Creative Commons
Dimensione 900.76 kB
Formato Adobe PDF
900.76 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/445299
Citazioni
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 3
Social impact