Purpose:The aim of this study was to assess the reliability of a 2D dark-blood phase-sensitive late gadolinium enhancement sequence (2D-DBPSLGE) compared with 2D phase-sensitive inversion recovery late gadolinium enhancement sequence (2D-BBPSLGE) in patients with ischemic cardiomyopathy (ICM).Materials and Methods:A total of 73 patients with a clinical history of ICM were prospectively enrolled. The following endpoints were evaluated: (a) comparison of image quality between 2D-BBPSLGEand 2D-DBPSLGEfor differentiation between blood pool-late gadolinium enhancement (LGE), remote myocardium-LGE, and blood pool-remote myocardium; (b) diagnostic accuracy of 2D-DBPSLGEcompared with gold standard 2D-BBPSLGEfor the evaluation of infarcted segments; (c) diagnostic accuracy of 2D-DBPSLGEfor the evaluation of microvascular obstruction (MVO); (d) comparison of transmurality index between 2D-BBPSLGEand 2D-DBPSLGE; (e) comparison of papillary muscle hyperenhancement between 2D-BBPSLGEand 2D-DBPSLGE; inter-reader agreement for depiction of hyperenhanced segments in both LGE sequences. Data were analyzed using paired t test, Wilcoxon test, and McNemar test, and η2coefficient and intercorrelation coefficient (ICC).Results:Image quality was superior for 2D-DBPSLGEfor differentiation of blood pool-LGE (P<0.001). 2D-DBPSLGE, compared with 2D-BBPSLGE, showed a sensitivity, specificity, positive predictive value, negative predictive value, and diagnostic accuracy of 96.93%, 99.89%, 99.71%, 98.78, and 99.04%, respectively. Concerning MVO detection, 2D-DBPSLGEshowed a sensitivity, specificity, positive predictive value, negative predictive value, and diagnostic accuracy of 66.67%, 100.00%, 100.00%, 80.95%, and 86.21%, respectively. 2D-DBPSLGEunderestimated the transmurality (P=0.007) and identified papillary muscle hyperenhancement (P<0.001). Both LGE sequences showed comparable interobserver agreement for the evaluation of infarcted areas (2D-BBPSLGE: ICC 0.99;2D-DBPSLGE: ICC 0.99).Conclusions:Compared with 2D-BBPSLGE, 2D-DBPSLGEsequences provide better differentiation between LGE and blood-pool, while underestimating LGE trasmurality and the presence of MVO.
Muscogiuri, G., Gatti, M., Dell'Aversana, S., Guaricci, A., Guglielmo, M., Baggiano, A., et al. (2020). Image Quality and Reliability of a Novel Dark-Blood Late Gadolinium Enhancement Sequence in Ischemic Cardiomyopathy. JOURNAL OF THORACIC IMAGING, 35(5), 326-333 [10.1097/RTI.0000000000000448].
Image Quality and Reliability of a Novel Dark-Blood Late Gadolinium Enhancement Sequence in Ischemic Cardiomyopathy
Muscogiuri G;
2020
Abstract
Purpose:The aim of this study was to assess the reliability of a 2D dark-blood phase-sensitive late gadolinium enhancement sequence (2D-DBPSLGE) compared with 2D phase-sensitive inversion recovery late gadolinium enhancement sequence (2D-BBPSLGE) in patients with ischemic cardiomyopathy (ICM).Materials and Methods:A total of 73 patients with a clinical history of ICM were prospectively enrolled. The following endpoints were evaluated: (a) comparison of image quality between 2D-BBPSLGEand 2D-DBPSLGEfor differentiation between blood pool-late gadolinium enhancement (LGE), remote myocardium-LGE, and blood pool-remote myocardium; (b) diagnostic accuracy of 2D-DBPSLGEcompared with gold standard 2D-BBPSLGEfor the evaluation of infarcted segments; (c) diagnostic accuracy of 2D-DBPSLGEfor the evaluation of microvascular obstruction (MVO); (d) comparison of transmurality index between 2D-BBPSLGEand 2D-DBPSLGE; (e) comparison of papillary muscle hyperenhancement between 2D-BBPSLGEand 2D-DBPSLGE; inter-reader agreement for depiction of hyperenhanced segments in both LGE sequences. Data were analyzed using paired t test, Wilcoxon test, and McNemar test, and η2coefficient and intercorrelation coefficient (ICC).Results:Image quality was superior for 2D-DBPSLGEfor differentiation of blood pool-LGE (P<0.001). 2D-DBPSLGE, compared with 2D-BBPSLGE, showed a sensitivity, specificity, positive predictive value, negative predictive value, and diagnostic accuracy of 96.93%, 99.89%, 99.71%, 98.78, and 99.04%, respectively. Concerning MVO detection, 2D-DBPSLGEshowed a sensitivity, specificity, positive predictive value, negative predictive value, and diagnostic accuracy of 66.67%, 100.00%, 100.00%, 80.95%, and 86.21%, respectively. 2D-DBPSLGEunderestimated the transmurality (P=0.007) and identified papillary muscle hyperenhancement (P<0.001). Both LGE sequences showed comparable interobserver agreement for the evaluation of infarcted areas (2D-BBPSLGE: ICC 0.99;2D-DBPSLGE: ICC 0.99).Conclusions:Compared with 2D-BBPSLGE, 2D-DBPSLGEsequences provide better differentiation between LGE and blood-pool, while underestimating LGE trasmurality and the presence of MVO.File | Dimensione | Formato | |
---|---|---|---|
muscogiuri2019_Dark blood LGE (1).pdf
Solo gestori archivio
Tipologia di allegato:
Publisher’s Version (Version of Record, VoR)
Dimensione
373.61 kB
Formato
Adobe PDF
|
373.61 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.