An improved method for the evaluation of glucose turnover rate in humans, using a prime-continuous infusion of [6,6-2H2]glucose, was developed. Deproteinization of plasma and conversion of glucose into the aldononitrile pentaacetate derivative are the only sample manipulations required prior to the gas chromatographic-mass spectrometric analysis. In six normal adults (prime = 5 mg kg-1; continuous infusion = 0.05 mg kg-1 min-1) the hepatic glucose output calculated at steady state by the procedure described here was 2.1 +/- 0.2 mg kg-1 min-1, the isotopic enrichment being determined with a coefficient of variation of ca. 2%. In six additional subjects, with half of the above-mentioned doses of labelled glucose, the mean hepatic glucose output was exactly the same (3.2% coefficient of variation for the isotopic enrichment measurement). The method described allows the hepatic glucose output to be precisely determined with savings both of time and of labelled glucose.
Magni, F., Monti, L., Brambilla, P., Poma, R., Pozza, G., Galli Kienle, M. (1992). Determination of plasma [6,6-2H2]glucose enrichment by a simple and accurate gas chromatographic-mass spectrometric method. JOURNAL OF CHROMATOGRAPHY, 573(1), 127-131.
Determination of plasma [6,6-2H2]glucose enrichment by a simple and accurate gas chromatographic-mass spectrometric method
MAGNI, FULVIO;Brambilla, P;
1992
Abstract
An improved method for the evaluation of glucose turnover rate in humans, using a prime-continuous infusion of [6,6-2H2]glucose, was developed. Deproteinization of plasma and conversion of glucose into the aldononitrile pentaacetate derivative are the only sample manipulations required prior to the gas chromatographic-mass spectrometric analysis. In six normal adults (prime = 5 mg kg-1; continuous infusion = 0.05 mg kg-1 min-1) the hepatic glucose output calculated at steady state by the procedure described here was 2.1 +/- 0.2 mg kg-1 min-1, the isotopic enrichment being determined with a coefficient of variation of ca. 2%. In six additional subjects, with half of the above-mentioned doses of labelled glucose, the mean hepatic glucose output was exactly the same (3.2% coefficient of variation for the isotopic enrichment measurement). The method described allows the hepatic glucose output to be precisely determined with savings both of time and of labelled glucose.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.