Seven arterial grafts implanted for a period varying from 8 months to 12 years and removed due to clinical failure were examined by means of an ISI-Super Mini SEM (Scanning Electron Microscope) at 7.5 Kv. Three specimens of dacron prostheses before implantation were also examined. The SEM findings are reported.

Biasi, G., Mingazzini, P., Sforza, V., Tanganelli, P., Ruberti, U., Weber, G. (1980). Scanning electron microscopic aspects of arterial dacron prostheses examined before and after long term implantation. SURGERY IN ITALY, 10(3), 227-232.

Scanning electron microscopic aspects of arterial dacron prostheses examined before and after long term implantation

BIASI, GIORGIO MARIA;MINGAZZINI, PAOLO;
1980

Abstract

Seven arterial grafts implanted for a period varying from 8 months to 12 years and removed due to clinical failure were examined by means of an ISI-Super Mini SEM (Scanning Electron Microscope) at 7.5 Kv. Three specimens of dacron prostheses before implantation were also examined. The SEM findings are reported.
Articolo in rivista - Articolo scientifico
Arterial Prostheses, Vascular Grafts, Biological Integration, Scanning Electron Microscopy
English
1980
10
3
227
232
open
Biasi, G., Mingazzini, P., Sforza, V., Tanganelli, P., Ruberti, U., Weber, G. (1980). Scanning electron microscopic aspects of arterial dacron prostheses examined before and after long term implantation. SURGERY IN ITALY, 10(3), 227-232.
File in questo prodotto:
File Dimensione Formato  
11.pdf

accesso aperto

Tipologia di allegato: Other attachments
Dimensione 3.89 MB
Formato Adobe PDF
3.89 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/34133
Citazioni
  • Scopus 0
  • ???jsp.display-item.citation.isi??? ND
Social impact