BACKGROUND: Endotracheal suctioning is mandatory to prevent complications caused by the retention of tracheal secretions. Endotracheal suctioning is often performed late, when patients show signs of respiratory and hemodynamic alterations. We conceived a prototype device that, when synchronized with the ventilator, automatically removes secretions collected below the endotracheal tube (ETT) cuff, thus avoiding endotracheal suctioning. The aim of our investigation was to assess the performance of this novel prototype in vitro. METHODS: Three studies were performed to examine the characteristics of the prototype. We tested device’s ability to generate an effective artificial cough flow (artificial cough maneuver) > 1 L/s by rapidly deflating the ETT cuff within the time of a sustained inflation (at 30 and at 40 cm H2O) (cough flow study). We also tested the prototype’s ability to remove the fluid positioned below the ETT cuff using saline dye (fluid removal study), and to prevent the aspiration of saline dye from above the ETT cuff during the deflation phase of the ETT cuff (aspiration study). The trachea model was positioned at 45° in the aspiration study, and horizontally in the other two studies. RESULTS: In the cough flow study, the prototype provided an effective artificial cough maneuver, with a mean ± SD of 1.78 ± 0.19 L/s (range, 1.42–2.14 L/s). The tracheal pressure after ETT cuff deflation never decreased below the PEEP level. In the fluid removal study, the prototype cleared the fluid from below the ETT cuff and the experimental trachea. No fluid was aspirated from the area above the ETT cuff toward the lower airways. CONCLUSIONS: We conceived an system capable of automatically expelling fluid from below the ETT cuff outside an experimental trachea by generating an artificial cough maneuver. This system may decrease the use of endotracheal suctioning and its complications. Future in vivo studies are needed to confirm this first in vitro evaluation.

Zanella, A., Florio, G., Rezoagli, E., Pastore, M., Cadringer, P., Biancolilli, O., et al. (2019). An artificial cough maneuver to remove secretions from below the endotracheal tube cuff. RESPIRATORY CARE, 64(4), 372-383 [10.4187/respcare.06076].

An artificial cough maneuver to remove secretions from below the endotracheal tube cuff

Zanella A.
Primo
;
Rezoagli E.;Pesenti A. M.
2019

Abstract

BACKGROUND: Endotracheal suctioning is mandatory to prevent complications caused by the retention of tracheal secretions. Endotracheal suctioning is often performed late, when patients show signs of respiratory and hemodynamic alterations. We conceived a prototype device that, when synchronized with the ventilator, automatically removes secretions collected below the endotracheal tube (ETT) cuff, thus avoiding endotracheal suctioning. The aim of our investigation was to assess the performance of this novel prototype in vitro. METHODS: Three studies were performed to examine the characteristics of the prototype. We tested device’s ability to generate an effective artificial cough flow (artificial cough maneuver) > 1 L/s by rapidly deflating the ETT cuff within the time of a sustained inflation (at 30 and at 40 cm H2O) (cough flow study). We also tested the prototype’s ability to remove the fluid positioned below the ETT cuff using saline dye (fluid removal study), and to prevent the aspiration of saline dye from above the ETT cuff during the deflation phase of the ETT cuff (aspiration study). The trachea model was positioned at 45° in the aspiration study, and horizontally in the other two studies. RESULTS: In the cough flow study, the prototype provided an effective artificial cough maneuver, with a mean ± SD of 1.78 ± 0.19 L/s (range, 1.42–2.14 L/s). The tracheal pressure after ETT cuff deflation never decreased below the PEEP level. In the fluid removal study, the prototype cleared the fluid from below the ETT cuff and the experimental trachea. No fluid was aspirated from the area above the ETT cuff toward the lower airways. CONCLUSIONS: We conceived an system capable of automatically expelling fluid from below the ETT cuff outside an experimental trachea by generating an artificial cough maneuver. This system may decrease the use of endotracheal suctioning and its complications. Future in vivo studies are needed to confirm this first in vitro evaluation.
Articolo in rivista - Articolo scientifico
Cough; In vitro; Intubation; Mechanical ventilation; Ventilator-associated pneumonia;
Cough; In vitro; Intubation; Mechanical ventilation; Ventilator-associated pneumonia; Cough; Equipment Design; Exhalation; Humans; Materials Testing; Pulmonary Elimination; Respiration, Artificial; Ventilators, Mechanical; Intubation, Intratracheal; Suction
English
2019
64
4
372
383
open
Zanella, A., Florio, G., Rezoagli, E., Pastore, M., Cadringer, P., Biancolilli, O., et al. (2019). An artificial cough maneuver to remove secretions from below the endotracheal tube cuff. RESPIRATORY CARE, 64(4), 372-383 [10.4187/respcare.06076].
File in questo prodotto:
File Dimensione Formato  
372.full.pdf

accesso aperto

Tipologia di allegato: Publisher’s Version (Version of Record, VoR)
Dimensione 1.12 MB
Formato Adobe PDF
1.12 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/287192
Citazioni
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 7
Social impact