In this paper, we deal with the problem of tracking a desired plasma glucose concentration by means of intravenous insulin administration, for Type 2 diabetic patients. A nonlinear time-delay model is used to describe the glucose-insulin regulatory system, according to which a model-based approach is exploited to design a semiglobal sampled-data dynamic output feedback controller. It is shown that emulation, by Euler approximation, of a proposed continuous-Time control law yields stabilization in the sample-And-hold sense of the glucose-insulin system. The glucose regulator makes use of only sampled glucose measurements. Theoretical results are preclinically validated through a virtual environment broadly accepted as a substitute to animal trials for the preclinical testing of control strategies in plasma glucose regulation. Numerical results are encouraging and pave the way to further clinical verifications.

Di Ferdinando, M., Pepe, P., Palumbo, P., Panunzi, S., De Gaetano, A. (2020). Semiglobal Sampled-Data Dynamic Output Feedback Controller for the Glucose-Insulin System. IEEE TRANSACTIONS ON CONTROL SYSTEMS TECHNOLOGY, 28(1), 16-32 [10.1109/TCST.2018.2881662].

Semiglobal Sampled-Data Dynamic Output Feedback Controller for the Glucose-Insulin System

Palumbo, P;
2020

Abstract

In this paper, we deal with the problem of tracking a desired plasma glucose concentration by means of intravenous insulin administration, for Type 2 diabetic patients. A nonlinear time-delay model is used to describe the glucose-insulin regulatory system, according to which a model-based approach is exploited to design a semiglobal sampled-data dynamic output feedback controller. It is shown that emulation, by Euler approximation, of a proposed continuous-Time control law yields stabilization in the sample-And-hold sense of the glucose-insulin system. The glucose regulator makes use of only sampled glucose measurements. Theoretical results are preclinically validated through a virtual environment broadly accepted as a substitute to animal trials for the preclinical testing of control strategies in plasma glucose regulation. Numerical results are encouraging and pave the way to further clinical verifications.
Articolo in rivista - Articolo scientifico
Glucose-insulin model; in silico validation.; nonlinear time-delay systems; sampled-data output feedback controllers; stabilization in the sample-and-hold sense
English
7-dic-2018
2020
28
1
16
32
8567971
open
Di Ferdinando, M., Pepe, P., Palumbo, P., Panunzi, S., De Gaetano, A. (2020). Semiglobal Sampled-Data Dynamic Output Feedback Controller for the Glucose-Insulin System. IEEE TRANSACTIONS ON CONTROL SYSTEMS TECHNOLOGY, 28(1), 16-32 [10.1109/TCST.2018.2881662].
File in questo prodotto:
File Dimensione Formato  
DiFerdinando_v3.pdf

accesso aperto

Descrizione: Versione post-print completa del lavoro
Tipologia di allegato: Author’s Accepted Manuscript, AAM (Post-print)
Dimensione 1.39 MB
Formato Adobe PDF
1.39 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/246875
Citazioni
  • Scopus 18
  • ???jsp.display-item.citation.isi??? 19
Social impact